Item type |
Multiple(1) |
公開日 |
2025-04-02 |
タイトル |
|
|
タイトル |
The Design and Characterization of Dielectric Microcalorimeters for X-ray Photon Detection |
|
言語 |
en |
作成者 |
菊池
星野
Nakayama
Yamasaki
前畑
Mitsuda
|
アクセス権 |
|
|
アクセス権 |
metadata only access |
|
アクセス権URI |
http://purl.org/coar/access_right/c_16ec |
内容記述 |
|
|
内容記述タイプ |
Abstract |
|
内容記述 |
The dielectric micro calorimeter (DMC) is a novel radiation detector utilizing a GHz resonator with dielectric thermometer (Sekiya et al. in J Low Temp Phys 167:435, 2012). The advantage of using a DMC is that the detection mechanism is based on a phonon mediation without Johnson noise and quasi-particle decay process. A large format array of DMCs can be easily multiplexed by a resonator circuit in the readout at GHz band width. We describe the design of a DMC as an X-ray photon counter. It is optimized to detect photon at 5.9 keV energy. We consider O-doped SrTiO (STO18) and Nb-doped KTaNbO (KTN) as a candidate of dielectric thermometer. Dielectric materials which have sensitivity (= d) are also ideal for our application. We check that both STO18 and KTN have at 100 mK. If we assume a DMC resonator operating at 100 mK, we need a Q value of a resonator to be 2000 for X-ray detection. |
|
言語 |
en |
日付 |
|
|
日付 |
2016-7 |
|
日付タイプ |
Issued |
Language |
|
|
言語 |
en |
資源タイプ |
|
|
資源タイプ識別子 |
http://purl.org/coar/resource_type/c_6501 |
|
資源タイプ |
journal article |
関連情報 |
|
|
関連タイプ |
isIdenticalTo |
|
|
識別子タイプ |
WOS |
|
|
関連識別子 |
WOS:000378689800040 |
関連情報 |
|
|
関連タイプ |
isIdenticalTo |
|
|
識別子タイプ |
SCOPUS |
|
|
関連識別子 |
84961794338 |
収録物識別子 |
|
|
収録物識別子タイプ |
ISSN |
|
収録物識別子 |
0022-2291 |
収録物名 |
|
|
収録物名 |
JOURNAL OF LOW TEMPERATURE PHYSICS |
巻 |
|
|
巻 |
184 |
号 |
|
|
号 |
1-2 |
ページ数 |
|
|
ページ数 |
7 |
開始ページ |
|
|
開始ページ |
250 |
開始ページ |
|
|
開始ページ |
256 |